
Measurement basics
R&S
®
NRX
47
User Manual 1178.5566.02 ─ 09
–
"Use Default"
Uses the sensor settings. For details, see the user manual of the sensor.
–
"Cancel"
Keeps the measurement settings and does not assign the sensor to the mea-
surement.
If conflicting settings occur when connecting a sensor, the R&S NRX shows where
the problematic setting is located. See
Chapter 5.5, "Settings conflict"
Example: Using different ports for the same sensor type
1. Connect an R&S
NRQ6 to port
A.
2. Perform a trace measurement.
3. Remove the R&S
NRQ6 and connect it to port
B.
The R&S NRX recognizes the sensor type and assigns the R&S
NRQ6 to the
same measurement.
Example: Using two sensors of the same type
1. Connect the first R&S
NRP-Z81 to port
A.
2. Connect the second R&S
NRP-Z81 to port
B.
3. Swap ports.
The R&S NRX distinguishes sensors of the same type due to their unique serial
number and assigns them to the same measurement as before.
Sensor assignment and memory