
Remote control commands
R&S
®
NRX
391
User Manual 1178.5566.02 ─ 09
DIAGnostic:INFO:OTIMe?
Queries the count of the built-in elapsed-time meter. The count is always output in
hours [h] and cannot be changed.
Usage:
Query only
TEST:DEVice:RESult?
.
Usage:
Query only
TEST:DEVice[:ALL]
[<argument>]
Performs tests for the keyboard, display and touch panel.
Parameters:
<argument>
The tests can be performed as single tests or as combined test.
"SubSystemGui:KeyboardTest"
Keyboard test
"SubSystemGui:DisplayTest"
Display test
"SubSystemGui:TouchTest"
Touch test
Example:
TEST:DEV "SubSystemGui:
KeyboardTest;DisplayTest;TouchTest";*OPC
Performs a combined test.
TEST:USB:STORage?
[<argument>]
Checks the connected memory stick.
Query parameters:
<argument>
0
Test passed
1
Test failed; no memory stick connected or memory stick does
not respond.
Usage:
Query only
TEST:SENSor<Sensor>?
Starts a selftest of the selected power sensor and returns the result. In contrast to
, this command returns detailed information that you can use for troubleshoot-
ing.
Running selftests