11-10
11. Specifications
Oxygen concentration controlling response time
The responded time of the oxygen concentration in
the delivered volume to change from a volume fraction
of 21 % to 90 % of the maximum settable oxygen
concentration:
When TV=500 mL, f=10 /min, I:E=1:2,
≤
50 s ;
When TV=150 mL, f=20 /min, I:E=1:2,
≤
100 s;
When TV=30 mL, f=30 /min, I:E=1:2,
≤
150 s.
11.6 Monitored Parameters
For all measured and computed variables that are
displayed or used for control, the low pass filtering and
moving average smoothing techniques is used.
Table 11-7.Dual-limb monitored parameters
Parameter
Range
Resolution
Accuracy
Paw
0~100 cmH
2
O
1 cmH
2
O
/
Ptrachea
0~100 cmH
2
O
1 cmH
2
O
/
Flow
-200~200 L/min
-99.9 to 99.9L/
min: 0.1L/min
-200 to -100/100
to 200L/min: 1L/
min
/
Volume
0~4000mL
1mL
/
P-peak
-10~100 cmH
2
O 1 cmH
2
O
± (2 cmH
2
O + 4 %
of actual reading )
P-plat.
-10~100 cmH
2
O
≥
10: 1 cmH
2
O
<10: 0.1 cmH
2
O
± (2 cmH
2
O + 4 %
of actual reading)
P-mean
-10~100 cmH
2
O
≥
10: 1 cmH
2
O
<10: 0.1 cmH
2
O
± (2 cmH
2
O + 4 %
of actual reading)
PEEP
-10~100 cmH
2
O
≥
10: 1cmH
2
O
<10: 0.1 cmH
2
O
± (2 cmH
2
O + 4 %
of actual reading)
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...