7-19
7. Alarms
ventilation in VCV mode normally.
2. Set the VT low alarm limit to be greater than the
current VTe.
3. Verify that the [
Low Tidal Volume
] alarm is activated.
7.12.11 Low Minute Volume
1. Make sure a demonstration lung is connected to
the ventilator and start ventilation in VCV mode
normally.
2. Set the MV low alarm limit to be greater than the
current MV.
3. Verify that the Low Tidal Volume alarm is activated.
7.13 Alarm Troubleshooting Table
The below table is a list of the alarm messages
displayed by the ventilator, along with their definitions
and suggested corrective actions. These corrective
actions are sequenced to correct the most probable
malfunction or to present the most efficient corrective
action first. The proposed actions, however, may not
always correct the particular problem.
This chapter lists physiological and technical alarm
messages.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...