11-3
11. Specifications
90
Max delivered pressure(cmH20)
Altitude(m)
80
70
60
50
40
30
20
0
1000
2000
3000
4000
5000
Figure 11-1.Altitude/Pressure changes
NOTE
•
The environment required for oxygen sensor is
listed as follows table.
Item
Temperature
Relative humidity
(non-condensing)
Barometric
pressure
Operating 0 to 50 ℃
0 to 99 %RH
0.6 to 2 bar
Storage
-20 to 50 ℃
Recommend 5
to 15 ℃
0 to 99 %RH
0.6 to 2 bar
The environment required for battery is listed as follows
table.
Item
Temperature
Relative humidity
(non-condensing)
Barometric
pressure
Operating
Charge: 0 to 45
℃
Discharge: -20
to 60 ℃
10 to 96 %RH
0 to 5000 m
Storage
-20 to 50 ℃
Recommend
-20 to 25 ℃
10 to 96 %RH
0 to 5000 m
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...