7-11
7. Alarms
Figure 7-4.Adjusting the alarm settings
7.8 Nurse Call
The ventilator provides nurse call function that enables
the ventilator to output nurse call signals to the
nurse call system when an alarm meeting the user set
requirements occurs.
The nurse call function is activated only when:
1. The nurse call function is switched on;
2. An alarm which meets the user set requirements
occurs;
3. The ventilator is not in AUDIO PAUSED status.
Follow the steps below to set nurse call:
1. Select [
Setup
] → [
Maintain
] → Enter user password
→ [
System
] → [
Nurse Call
].
2. Select [
Switch
] and toggle between [
ON
] and [
OFF
].
[
ON
]: to switch on the nurse call function.
[
OFF
]: to switch the nurse call function off.
3. Set [
Signal type
].
[
Pulse
]: indicates that the nurse call signals
outputted are pulse signals lasting for one second.
When multiple alarms occur simultaneously, only
one pulse signal is outputted. If a new alarm occurs
when the ongoing alarm is not cleared yet, a new
pulse signal will be outputted.
245 - 600
XXXXXXXXXX
XXXXX
XXXXX
XXXXX
XXXXX
XXXXX
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...