F. Neonatal Ventilation
F-6
When the neonatal flow sensor returns to normal, the
ventilator will switch back to the original ventilation
mode automatically.
F.6 Neonatal Flow Sensor calibration
Please perform calibration of the neonatal flow sensor
when the measured value has a great deviation. Refer
to
9.5 Flo
w
s
en
s
o
r
cali
br
ation
.
NOTE
•
Please pay attention to the connection direction of
the sensor during calibration.
•
First the forward ventilation, then the reverse
ventilation.
F.7 Control Settings
Timing trigger, pressure trigger and flow trigger are
used for initiating and terminating the inspiratory
phase in ventilator, but different mode use different
means.
NOTE
•
The unlisted parameters are the same as the
corresponding parameters of adult.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...