F. Neonatal Ventilation
F-4
F.2 Installing Neonatal Tubing
Refer to
3.2 In
s
tallin
g
the Patient
Tub
in
g
.The use of a
humidifier is recommended when installing neonatal
tubing.
1. Mount the filters onto the inspiratory and expiratory
ports.
2. Connect the inspiratory filter to the humidifier inlet
via the neonatal tube.
3. Connect the humidifier outlet to the water trap via
the neonatal tubing. Then connect the water trap to
the Y piece via the neonatal tubing.
4. Connect the expiratory filter to the water trap via
the neonatal tubing. Then connect the water trap to
the Y piece via the neonatal tubing.
5. Connect the neonatal flow sensor tubing to the
ventilator.
6. Connect the small end of the neonatal flow sensor
to the Y piece, and the large end to the neonatal
test lung.
7. Place the neonatal tubing onto the support arm
hook.
1-Y piece of the neonatal tubing
2-Neonatal flow sensor
3-Neonatal test lung
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...