7. Alarms
Item Alarm
Priority
Definition
Action needed
53
Inspiratory
time too long
Low
Spontaneous b
reathing fails t
o satisfy the ex
p. sensitivity al
l the time so th
at expiration is
unable to end.
Disconnect the
patient. Restart the
ventilator.
54
Patient circuit
partially
occluded
High
Exp. limb airway
obstructed
Check the patient.
Check the expiratory
limb for occlusion.
Check the expiratory
valve membrane and
cover.
Check the Flow
Sensor tubes for
occlusion.
Adjust breath timing
controls to increase
the expiratory time.
55
Performance
limited by
high altitude
High
The altitude is
too high.
Adjust breath
controls.
56
Please
calibrate
pressure
sensor
High
Pressure sensor
is not calibrated
Perform pressure se
nsor calibration.
57
Please change
the blower
Low
Blower total ru
nning time exc
eeds the factor
y confirmed ma
ximum faultless
running time.
Contact your
service personnel.
58
Please check
patient flow
sensor
High
Patient flow
sensor is faulty.
Calibrate the
Patient flow
sensor.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...