7-23
7. Alarms
Item Alarm
Priority
Definition
Action needed
5
Battery
Temperature
High
Syst Maybe
Down
High
Battery
temperature is
too high during
discharge.
The system may
be down.
Connect to the
external power
supply.
Remove the
ventilator from the
sun or other heat
source.
6
Fan Failure
Medium
The cooling
fan is
malfunctioning.
Disconnect the
ventilator from the
patient. Have the
ventilator serviced.
7
HEPA
Replacement
Suggested
Low
The air inlet
HEPA filter
shows increased
resistance.
Replace the HEPA
filter.
8
High Blower
Temperature
High
The blower
temperature is
too high.
The ventilation will
be stopped. Contact
service providers.
9
High PEEP
High
Monitored PEEP
exceeds PEEP +
5 cmH2O for 3
fully mechanical
ventilation cycle.
Check the patient.
Confirm ventilator
and alarm settings
are appropriate.
10
High
Sustained
Airway
Pressure
High
The measured
airway pressure
is greater than
or equal to
the setting
PEEP+15cmH2O
for continuous
15s.
Check the patient.
Check the ventilation
parameter setup.
Check the breathing
tubes for occlusion.
11
Inspiratory
Gas
Temperature
High
High
The gas
temperature
exceeds 45℃ .
Disconnect the
patient. Restart the
ventilator.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...