6-26
6. Start Ventilation
6.6 Tube Resistance Compensation
Synchronized tube resistance compensation (STRC)
stands for the function of automatic tube resistance
compensation. By selecting appropriate endotracheal
(ET) tube or tracheostomy (Trach) tube of different
diameters for the user, the ventilator can adjust gas
delivery pressure automatically, so that the pressure at
the end of the tube is consistent with the ventilator’s
pressure setting value as much as possible.
3
10
1.70
OFF
XXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXX
XXXXXXXXXXX
XXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
Figure 6-20.STRC settings
1. Select [
Setup
] → [
Vent
] → [
STRC
].
2. Set the items below on the accessed screen:
◆ [
Tube Compensation
]: disable, ET Tube or Trach
Tube.
◆ [
TubeID
]: ET tube diameter.
◆ [
Compensation
]: percentage of STRC.
◆ [
Expiratory Compensation
]: enable or disable
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...