9-16
9. Maintenance
mA)on the BF type applied parts; that the maximum
leakage current in the middle four tests is not higher
than 50 μA(0.05 mA)on the CF type applied parts
and not higher than 500 uA(0.5 mA)on the BF type
applied parts; that the maximum leakage current in the
last two tests is not higher than 50 μA(0.05 mA)on
the CF type applied parts and not higher than 5000 uA
(5 mA)on the BF type applied parts.
NOTE
•
Ensure the safety analyzer is authorized by
certificate organizations (UL, CSA, or AAMI
etc.). Follow the instructions of the analyzer
manufacturer.