6-6
6. Start Ventilation
6.3 Select Patient
Open the patient setting menu in standby and select
the patient information:
◆ If selecting [
New Patient
], please set [
Patient
Category
], [
Gender
], [
Height
]/[
IBW
] and [
Ventilation
Type
] in the accessed [
New Patient
] menu.
◆ If selecting [
Last Patient
], only the [
Ventilation
Type
] can be set.
◆ Upon alteration of [
Gender
], [
Height
] or [
IBW
], the
settings of [
VT
], [
VTapnea
], [
f
] and [
fapnea
] will
change accordingly, as well as tidal volume high
alarm limit, tidal volume low alarm limit, minute
ventilation high alarm limit and minute ventilation
low alarm limit.
◆ The [
Patient 1
]/[
Patient 2
]/[
Patient 3
] are the user
preset for ventilation settings, you can customize
them for different patients.
NOTE
•
To prevent the possible patient injury, make sure
the ventilator is set up for the appropriate patient
type with the appropriate breathing circuit parts.
After the System Check pass, the ventilator
configuration is displayed on the Standby Windows.
You can set up the patient type, patient circuit, and
components of the breathing system for new and
resuming patients.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...