7. Alarms
Parameter
Range
Resolution
Factory
default
FiO
2
High Limit
(Low-pressure
oxygen supply)
20 ~ 100%
1%
100%
FiO
2
Low Limit
(Low-pressure
oxygen supply)
18 ~ 98%
1%
21%
Inspiratory
Pressure High
Limit
2 ~ 50 cmH2O
1 cmH
2
O
50 cmH
2
O
Inspiratory
Pressure Low
Limit
Off, 1 ~ 48 cmH2O
1 cmH
2
O
off
VT High Limit
10 ~ 2500 mL
5mL
2500
VT Low Limit
Off, 5 ~ 2490 mL;
5mL
OFF
f High Limit
(Low Limit +2) to 90 /min
1 /min
30
f Low Limit
1 to (High Limit -2) /min
1 /min
10
LIPT (Low
Inspiratory
Pressure Delay
Time)
5 ~ 60 s
1s
20s
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...