Remote Control - Commands
R&S
®
NRP2
212
User Manual 1173.9157.02 ─ 03
<expected_value>
Value range: float_value | DEF
According to the SCPI-1999.0 standard, a value in the order of the expected
measured value can be specified. This allows the device to set an adequate mea-
surement range. This is not required in the R&S
NRP2 because the measurement
is performed simultaneously in all measurement ranges. Any value can therefore
be specified for the expected_value parameter, especially the DEFault value. The
value is always ignored.
<no_slots>
Value range: depending on sensor
Default value: depending on sensor
Unit: -
Number of timeslots to be measured in the Timeslot mode. Corresponds to com-
mand:
<resolution>
Value range: 1 | 0.1 | 0.01 | 0.001 |
1 | 2 | 3 | 4
Default value: 3 (0.01)
Specifies the limit up to which the measurement result should be free of noise.
The R&S
NRP2 tries to set an adequate window for the averaging filter. This does
not affect the accuracy of the measured data. The window can only be set if auto-
matic setting of the filter window has not been disabled by the user
(
SENSe:AVERage:COUNt:AUTO OFF
). Since the automatic setting is activated
by
CONFigure
and
MEASure?
as a default, the user normally does not need to
alter this setting. Corresponds to commands:
Example:
If
<resolution>
is set to 3 or 0.01, three significant points are assigned to the
value when it is displayed on a linear scale. When a log scale is used, 2 places
are displayed after the decimal point.
<scope_size>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
The number of test points on the time axis. Corresponds to command:
<source_list>
Value range: (
@
n) | (
@
n),(
@
m)
Default value: (
@
1) or (
@
1),(
@
2)
n and m may assume the values 1, 2, 3 or 4.
With this list, sensors are assigned to the primary and possibly to the secondary
channel of a calculate block. The number for each sensor is preceded by the
character
@
and the entire expression is enclosed in parentheses.
Examples:
(
@
1)is used for sensor
A.
The values (
@
3),(
@
2) are set for <source_list>. Sensor C is assigned to the pri-
mary channel and sensor
B to the secondary channel. The calculate blocks proc-
ess the primary channel together with the secondary depending on the desired
function. If
:RATio
is selected as the calculation function, the result from the
calculate block is the ratio of the values measured by sensor
C and sensor
B.
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...