Getting Started
R&S
®
NRP2
32
User Manual 1173.9157.02 ─ 03
All sensor-related settings are made in the "Sensor" menu. It determines type and
details of data acquisition.
2. Press the left side of the "Offset…" softkey.
The "Offset" dialog box opens. Offsets in dB for compensating external signal losses
or gains, for example by attenuator or amplifiers, can be set in this dialog box.
Fig. 2-8: Offset dialog box
3. Activate the entry for the global offset.
Positive values are for correcting losses and negative values for gains.
4. If you have an attenuator at hand, enter its value and connect the attenuator between
the sensor and the signal source, otherwise just enter 10
dB.
5. Confirm the entry with
↵
MENU.
6. Activate the global offset correction by pressing the right side of the associated soft-
key.
Fig. 2-9: Offset dialog box / Global activated
Tip:
Factors which are not, or only minimally, depending on frequency can be cor-
rected in this way.
7. Press the
↵
MENU hardkey twice to exit the dialog box and the menu.
If you have connected an attenuator and entered its attenuation, the R&S
NRP2 will
display about the same value as before.
Measuring average power (Cont Av mode)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...