Compatibility Information
R&S
®
NRP2
401
User Manual 1173.9157.02 ─ 03
D Compatibility Information
This section contains information regarding the compatibility of the R&S
NRP2 remote
control commands with those of the "E4418B" and "E4419B" power meters from Agilent.
The R&S
NRP2 is largely downward-compatible to the above-mentioned instruments but
offers additional capabilities.
E4418B/E4419B command
Compatibility information
CONFigure?
Returns the current configuration. The R&S
NRP2 returns the last set-
ting made with
CONFigure
, as specified by the SCPI standard.
CALCulate:GAIN
SENS:CORR:OFFS:STAT
CALCulate:PHOLd
Is implemented in the R&S
NRP2 and has the same effect as
CALC:EXTR:RES
.
CALibration
:ECONtrol
:RCALibration
:RFACtor
Is not implemented in the R&S
NRP2.
DISPlay:ENABle[?]
Is implemented in theR&S
NRP2 and has the same effect as
DISPlay:ILLUmination
.
DISPlay:SCREen:FORMat
[:NUMeric[1]|2]
Is not implemented in the R&S
NRP2.
INITiate
:CONTinuous:SEQUence[1]|2
:IMMediate:SEQUence[1]|2
The
:SEQUence
node is not implemented in the R&S
NRP2.
MEMory:TABLe:GAIN
Agilent specifies the values without a unit; the unit PCT (percent) is
implicitly assumed. The units DB (dB), DPCT (
Δ
%) and O (unit 1) can
be specified in the R&S
NRP2. The default unit is DB.
OUTPut:TRIGger:STATe
Is not implemented in the R&S
NRP2.
SENSe
:AVERage:COUNt:AUTO:
SDETect
:DETector:FUNCtion
:MRATe
:SPEed
:V2P
Is not implemented in the R&S
NRP2.
SENSe[1|2]:CORRection
:CSET[1]|CSET2[:SELect]
and
SENSe[1|2]:CORRection
:CSET[1]|CSET2:STATe
The R&S
NRP2 does not recognize any tables with frequency-depend-
ent calibration values since the sensors have to be calibrated only once
in production. For this reason, there is no compliance with
CSET1
.
Tables with frequency-dependent offset correction values can be
defined and addressed via
:CSET2
(Agilent and R&S
NRP2)
or
:FDOTable
(R&S
NRP2 only).
SENSe[1|2]:CORRection:
GAIN[1]?
Has the same effect as
SENSe[1..2]:CORRection:CFACtor?
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...