Remote Control - Commands
R&S
®
NRP2
195
User Manual 1173.9157.02 ─ 03
Since
*OPC?
waits until all previous commands are executed, "
1
" is returned in all cases.
*OPC?
basically functions like the
*WAI
command, but *
*WAI
does not return a response.
*OPC?
is preferred to
*WAI
because with
*OPC?
, the execution of commands can be
queried from a controller program before new commands are sent. This avoids an over-
flow of the input queue when too many commands are sent that cannot be executed.
Unlike
*WAI
,
*OPC?
must be sent at the end of a program message.
*OPT?
OPT
ion identification query returns a list of the options installed. The options are sepa-
rated from each other by means of commas.
Option
Response string
R&S NRP-B1
"NRP-B1: Test generator installed"
R&S NRP-B2
"NRP-B2: Second channel interface installed"
R&S NRP-B5
"NRP-B5: Four channel interface installed"
R&S NRP-B6
"NRP-B6: Sensor A/B rear socket"
Example:
*OPT?
Device response:
"NRP-B1: Test Generator installed","NRP-B2:
Second Channel Interface installed","NRP-B5:
Four Channel Interface installed"
Usage:
Query only
*PRE
<value>
Setting command:
*PRE
P
arallel poll
R
egister
E
nable sets the parallel poll enable register to the specified value.
Query:
*PRE?
The query form
*PRE?
returns the contents of the parallel poll enable register in decimal
form.
Parameters:
<Value>
0 to 255
*PSC
<function>
Setting command:
*PSC
P
ower on
S
tatus
C
lear determines whether the contents of the ENABle registers are
preserved or reset during power-up.
Query:
*PSC?
Common Commands to IEEE 488.2
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...