Error Messages and Troubleshooting
R&S
®
NRP2
338
User Manual 1173.9157.02 ─ 03
Error code
Error text returned upon queue query
Explanations
-168
"Block data not allowed"
The command contains a legal block data element at a point where it is not allowed.
-171
"Invalid expression"
The command contains an invalid mathematical expression data element.
-178
"Expression data not allowed"
The command contains mathematical expression data at a point where they are not allowed.
8.2.3 Execution Error
Execution errors; cause bit 4 in the ESR register to be set
Error code
Error text returned upon queue query
Explanations
-200
"Execution error"
An error occurred upon command execution.
-203
"Command protected"
An attempt was made to execute a protected command.
-210
"Trigger error"
Error on triggering the device.
-211
"Trigger ignored"
A
*TRG
or a triggering signal was ignored.
-213
"Init ignored"
A request for a measurement initialization was ignored as another measurement was already
in progress.
-214
"Trigger deadlock"
A measurement could not be started or a result query could not be processed since the device
would otherwise enter a deadlock state. This occurs under the following conditions:
(1)
TRIG:SOUR BUS;:INIT;:FETCh?:
A
*TRG
or
TRIG:IMM
would be required for triggering
but these commands cannot be executed since
FETCh?
blocks the entry. But
FETCh?
waits
until the measured values are available.
(2)
FETCh?
was called but no valid result was available and the trigger system of the sensor
is
IDLE. FETCh?
blocks the entry so that a measurement cannot be started with INIT either.
-220
"Parameter error"
The command contains a faulty or invalid parameter.
-221
"Settings conflict"
There is a setting conflict between two or more parameters.
-222
"Data out of range"
A program data element is outside the legal range as defined by the device.
-223
"Too much data"
A valid parameter is received but it contains more data than the device can handle.
SCPI-Specific Error Messages
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...