Programming Measurements
R&S
®
NRP2
380
User Manual 1173.9157.02 ─ 03
Trace measurement:
MEAS:XTIMe? (<scope_size>),<capture_time>
Measurement in the ContAv mode with data buffering:
MEAS:ARRay?
<tslot_width>,<no_slots>,<start_exclude>,<stop_exclude>
In the Timeslot mode, the trigger source is set to
EXTernal
. In the "BurstAv" mode, it is
not modified but ignored. Triggering in response to the signal is performed as if
TRIGger:SOURce INTernal
were set.
These commands have the specified parameters, whose meanings are explained in the
operating manual. In addition, parameters
DEF
,
<resolution>
,
<source_list>
can
be optionally specified as for the
MEAS?
command.
Example:
MEAS:TSLot? 577 us, 8, 18 us, 18 us
A timeslot measurement of a GSM signal is performed (8 timeslots with a length of
577
µs each). 18
µs at the beginning and the end of the timeslots are ignored. The mea-
surement is carried out with sensor 1 (default). The instrument is triggered by an external
signal, which must be routed to the R&S
NRP2 rear panel via the trigger input. The mea-
surement result is a list of 8 measured values separated by a comma for the timeslots.
Example:
MEAS:BURSt? 5 us, 10 us, 0 us, DEF, 3, (@2)
Sensor 2 measures in the BurstAv mode. Triggering (in response to a rising edge) is
internal. At the beginning of the power pulse, 10
µs are ignored to avoid overshoots in
the signal, which could distort the measurement result. If the power falls below the trigger
level but does not exceed 5
µs (dropout time) the measurement will not be stopped. The
measurement result is noise-free within the first 3 places and it is measured with sensor
2 ((@2)).
Example:
MEAS:XTIMe? (256), 577 us
Within the next 577
µs, 256 measured values are recorded and displayed with respect
to time in te same manner as with an oscilloscope in order to display the power charac-
teristics. There is no triggering (
TRIG:SOUR IMM
); the measurements are started after
the command is received from the R&S
NRP2. After a partial measurement has been
completed, the next one is started. If the trigger source setting is to be changed,
MEAS:XTIMe?
must be replaced by the
CONF:XTIMe
and
READ:XTIMe?
commands,
between which user-defined settings can be performed. Dividing
MEAS?
into the
CONF
and
READ?
commands is described further down. The measurement result is a list of 256
measured values separated by a comma.
Measuring with high-level commands
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...