Remote Control - Commands
R&S
®
NRP2
262
User Manual 1173.9157.02 ─ 03
Parameters:
<mode>
NONE
No additional values are measured.
MINMax
By averaging the measured values in the sensors, extreme values
are lost. Using
[SENSe<[1]..4>]:AUXiliary MINMax
, a
sensor can be assigned to transmit to the R&S
NRP2 the minima
and maxima that occurred in the Trace mode together with the
measured value. If
CALCulate:FEED "POWer:PEAK ON
SWEep<[1]..4>"
is set, peak (instead of average) values are
used for calculating.
RNDMax
In contrast to
MINMax
, instead of the Min value the value of a ran-
domly selected sample is returned. All evaluations occur using
these values instead of the average values.
*RST:
depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt
<filter_length>
Sets the filter length i.e. the number of readings to be averaged for one measured value..
The higher the count the lower the noise and the longer it takes to obtain a measured
value.
Parameters:
<filter_length>
int_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: -
[SENSe<[1]..4>]:AVERage:COUNt:AUTO
<state>
Can be used to automatically determine a value for the filter legth
[SENSe<[1]..4>]:AVERage:COUNt
. If the automatic switchover is activated with the
ON parameter, the sensor always defines a suitable filter length which can also be quer-
ied via
SENS<[1]..4>:AVER:COUN?
.
SENSe (Sensor Commands)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...