Fig. 2-29: Function & Unit dialog box / unit
The ratio of two powers is dimensionless – this is why only "dB", "
Δ
%" and "1" are
available in the unit list. The symbol "
Δ
%" represents the difference between the
power values in W, given in % (0
% means the powers in both channels are equal).
The symbol "1" represents the quotient of the power values in W, i.e. not the log of a
ratio.
Fig. 2-30: Result windows / sensor A and B
2.7 Graphically representing power versus time
In the "Trace & Statistics" mode, power versus time can be represented graphically as
with an oscilloscope. If you wish to carry out measurements on signals with modulated
power envelope (in the Burst or Timeslot mode), initially displaying the signal in the Trace
mode is always recommended. Stable triggering by the R&S
NRP2 can more easily be
accomplished in this mode.
Setting the "Trace & Statistics" mode.
1. Connect a power sensor of the R&S
NRP-Z1x, R&S
NRP-Z2x or R&S
NRP-Z8x family
to connector A of the R&S
NRP2.
2. Apply an amplitude-modulated or pulsed signal having a modulation frequency or
pulse frequency of approx. 1
kHz. The maximum level should be between –10
dBm
and +10
dBm.
A pulse with 250
μ
s width, a period of 1ms, and 4dBm is used in the following example.
Graphically representing power versus time
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...