Instrument Functions
R&S
®
NRP2
146
User Manual 1173.9157.02 ─ 03
4.7.1.2
GPIB settings
Fig. 4-74: Remote Interfaces / GPIB
For configuring the R&S
NRP2 for remote control using the GPIB interface, just enter its
GPIB address.
GPIB Address
Selects the address in the range 0 to 30 (factory preset is 20). The set GPIB address is
unchanged after a reset of the instrument settings.
SCPI command:
Command Set
Selects the instrument type for emulation, see
4.7.1.3
USB settings
Remote control of the R&S
NRP2 via the USB is performed according to the USBTMC
standard. For this purpose, a VISA library that supports remote control via the USB is
required by the remote-control computer.
Fig. 4-75: Remote Interfaces / USB
VISA Resource Identifier
Displays the VISA address string of the instrument. This identifier is automatically
assigned for remote control via USB, therefore the address cannot be configured for
remote control.
System settings
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...