Instrument Functions
R&S
®
NRP2
106
User Manual 1173.9157.02 ─ 03
Fig. 4-40: Setting the Unit
Only the units which are compatible with the selected measurement function are provided
for selection. I.e. absolute power values are displayed in W, dBm or dB
μ
V , and relative
values in 1, dB or
Δ
%, respectively.
The following tables show how the R&S
NRP2 calculates the measurement results for
the different units either in absolute notation or relative to a reference value, see
ter 4.4.4, "Relative measurement"
Table 4-3: Formulas for absolute single channel power measurement
Function
Unit
W
dBm
dB
μ
V
Single
P
mW
1
lg
10
P
V
1
lg
20
Z
P
Table 4-4: Formulas for power ratio measurement
Function
Unit
W
dBm
dB
μ
V
Ratio (A,B)
B
A
P
P
lg
10
1
100
B
A
P
P
B
A
P
P
SWR (A,B)
–
–
A
B
A
B
P
P
P
P
1
1
RCoeff (A,B)
–
–
A
B
P
P
RLoss (A,B)
B
A
P
P
lg
10
–
–
Configuring measurements
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...