Error Messages and Troubleshooting
R&S
®
NRP2
339
User Manual 1173.9157.02 ─ 03
Error code
Error text returned upon queue query
Explanations
-224
"Illegal parameter value"
The parameter value is invalid.
-225
Out of memory
The device has insufficient memory to perform the required operation.
-226
"Lists not same length"
A list of offset values and frequencies has not the same length as an associated list of frequency
and offset values.
-230
"Data corrupt or stale"
The data are incomplete or invalid.
-240
"Hardware error"
The command cannot be executed because of a hardware problem in the device.
-241
"Hardware missing"
The command cannot be executed because of missing device hardware.
8.2.4 Device-Specific Errors
Device-specific error; causes bit 5 to be set in the ESR register
Error code
Error text returned upon queue query
Explanations
-300
"Device-specific error"
Generic device-dependent error that cannot be defined more precisely.
-310
"System error"
Indicates that some system error has occurred. Please contact the R&S service center.
-311
"Memory error"
An error was detected in the device's memory.
-314
"Save/recall memory lost"
Indicates that the nonvolatile data saved by the
*SAV?
command has been lost.
-315
"Configuration memory lost"
Indicates that nonvolatile configuration data saved by the device has been lost.
-330
"Self-test failed"
The self-test could not be executed.
-341
"Zeroing failed"
The zeroing of a sensor (
CALibration:ZERO:AUTO ONCE
) has failed. The most frequent
cause for this error is a signal that is applied to the sensor input and prevents zeroing.
SCPI-Specific Error Messages
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...