"Reset" will set the current power result as new reference value. After the reset, rel-
ative measurement is automatically active again.
If the power has not changed in the meantime, the indicated value is close to 0
dB.
Fig. 2-13: Result window / Reset relative
5. If an attenuator is still connected, remove it now and connect the sensor directly to
the signal source.
The displayed value should now equal the offset that has been entered.
Checking the settings with full height window
1. Select and open the "Windows" menu.
The "Windows" menu contains all the functions required for opening, closing and
configuring windows.
Fig. 2-14: Window menu / Digital
2. Check if window
1 has been selected.
3. Press the "Expand" softkey.
Window
1 now expands to full size below the menu bar and displays all relevant
settings:
- measurement modes of contributing channels (A (ContAV))
- Trigger mode (auto)
- averaging number and mode (32 Auto)
- Offset (10 dB)
- S parameter device (none)
- duty cycle correction (none)
Measuring average power (Cont Av mode)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...