Remote Control Fundamentals
R&S
®
NRP2
168
User Manual 1173.9157.02 ─ 03
keyboard including ESC/LOCAL has been disabled via the
LLO
command. In this case,
returning to local mode is only possible with the remote control command
GTL
or by a
power-on reset.
5.1.2 Measurement Modes
To simplify operation in the manual mode, preset values are used in some cases for the
trigger system (
TRIGger
command) and the timing of the measurement
(
SENSe[1..4]:TIMing:EXCLude:STARt
and
:STOP
) when a measurement mode
such as "ContAv", "Burst", "Timeslot" or "Trace" is selected, see
,
settings for manual selection of a measurement mode
. This is not the case in the remote
control mode. Here the trigger system and the timing can be configured by the user as
required.
5.1.3 Calculate Blocks
The R&S
NRP2 manages 8 internal calculate blocks. Each calculate block has two input
channels (primary and secondary channel) to which one of up to 4 power sensors can
be assigned. The channels can be processed together by means of a selectable function.
The result is available at the output of the calculate block. In the manual mode, the mea-
surement results of the first four calculate blocks (
chapter 6.5, "CALCulate (Configuration
, on page 221) are shown in windows 1 to 4, while 8 calculate blocks
are available in the remote control mode.
5.2 Connecting a Controller to the Base Unit
5.2.1 Settings in Base Unit
Before a link can be established between the controller and the R&S
NRP2, the
R&S
NRP2 must be assigned a corresponding address.
In the manual control mode, the address is set in the "System" > "Remote" dialog box,
see
chapter 4.7.1, "Configuring the remote control interfaces"
In remote control mode, the
SYSTem:COMMunication:..
commands serve this pur-
pose.
GPIB configuration (IEC/IEEE bus):
1. Select "GPIB" as the active interface.
2. Enter the GPIB address. Any value between 0 and 30 is permissible.
USB configuration:
► Select "USB" as the active interface.
Connecting a Controller to the Base Unit
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...