Remote Control Fundamentals
R&S
®
NRP2
182
User Manual 1173.9157.02 ─ 03
Table 5-2: Synchronization with *OPC, *OPC? and *WAI
Command
Effect
Programming of controller
*OPC
Sets the Operation Complete bit in the ESR
after all commands sent prior to *OPC have
been processed.
Set bit 0 in the ESE of the R&S
NRP2.
Set bit 5 in the SRE of the R&S
NRP2.
Wait for service request (SRQ).
*OPC?
Writes 1 in the output buffer after all com-
mands sent prior to
*OPC?
have been pro-
cessed.
IEC/IEEE bus: the R&S
NRP2 is addressed
as a talker.
*WAI
Commands can be sent but they are only exe-
cuted after completion of all commands sent
prior to
*WAI
.
Dividing commands timewise before/after
*WAI
serializes the command sequence.
*WAI
is sent directly after the command that
should be completed before other com-
mands can be executed.
5.9 Automatic Settings for Manual Selection of a Measure-
ment Mode
When the measurement mode of a sensor is manually selected ("Sensor"
→
"Mode"),
automatic settings are made for the trigger system and timing, which cannot be changed
by manual control. This considerably simplifies operation. These settings are not changed
when the measurement mode is selected by remote control (
SENSe:FUNCtion
) and
have to be set to the desired value by appropriate SCPI commands.
The following table lists the respective setting capabilities and the values that are auto-
matically set in manual control.
Table 5-3: Automatic settings for manual selection of a measurement mode
SCPI command
Value
Remark
Mode Cont Avg
TRIGger[1..4]:SOURce
IMMediate
Cannot be modified in manual mode.
TRIGger[1..4]:SLOPe
-
Unchanged, is irrelevant.
TRIGger[1..4]:COUNt
1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
-
Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff
1 NS
Cannot be modified in manual mode.
TRIGger[1..4]:LEVel
-
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
-
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP
-
Unchanged, is irrelevant.
INITiate[1..4]:CONTinuous
ON
Cannot be modified in manual mode.
Mode Burst Av
TRIGger[1..4]:SOURce
-
Unchanged, is irrelevant.
Automatic Settings for Manual Selection of a Measurement Mode
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...