Instrument Functions
R&S
®
NRP2
88
User Manual 1173.9157.02 ─ 03
Dropout ← Trigger dialog box (2)
Sets the dropout time. The dropout time prevents the trigger system from being activated
by signal excursions prior to the trigger event targeted. The parameter is set to a value
that is slightly higher than the maximum duration of power drops that are supposed to
execute triggering.
In contrast to the "Holdoff" parameter which is referenced to the active slope (trigger
event), the dropout time takes effect if the passive slope of the trigger signal crosses the
trigger threshold.
SCPI command:
Page ← Trigger dialog box (2)
Switches to the next page of the dialog box.
Trigger dialog box (3)
The third page of the trigger dialog box covers the hysteresis and holdoff parameters.
Fig. 4-22: Trigger dialog box, page 3
A | B | C | D ← Trigger dialog box (3)
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Hysteresis ← Trigger dialog box (3)
Sets the trigger hysteresis. The trigger hysteresis prevents the trigger system from being
activated too early if the trigger threshold is slightly underranged or exceeded. Before
triggering can occur again, the signal level must be lower (higher) than the trigger level
minus (plus) hysteresis.
Modifications of the trigger hysteresis are effective immediately (see
SCPI command:
Holdoff ← Trigger dialog box (3)
Sets the holdoff time. The holdoff time starts with the trigger event. During holdoff time
other trigger events are suppressed.
Data acquisition and parameters
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...