Remote Control - Commands
R&S
®
NRP2
328
User Manual 1173.9157.02 ─ 03
TRIGger<[1]..4>:DELay
<delay>
TRIGger:ALL:DELay
<delay>
Defines the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
TRIGger<[1]..4>:DELay:AUTO
<state>
TRIGger:ALL:DELay:AUTO
<state>
TRIG:DEL:AUTO ON
makes certain that a measurement is not started until the sensor
has settled. To this end, the delay value is automatically determined. This is especially
important when thermal sensors are used. The automatically determined delay is ignored
if a longer period is set with
TRIG<[1]..4>]:DEL
.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
TRIGger<[1]..4>:DTIMe
<dtime>
TRIGger:ALL:DTIMe
<dtime>
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted signal
components can thus be suppressed. This setting parameter is used with pulsed GSM
signals, for example, if only specific timeslots are to be measured.
The figure below shows the importance of the drop-out time. If the timeslots BF and DH
are not to be measured, the drop-out time is set to a higher value than the period EB
(GD). In this case, the trigger events B and D are ignored.
Trigger level
Power
Time
Drop-Out
Time
Trigger hysteresis
A
B
C
D
E
F
G
H
Drop-Out
Time
Fig. 6-20: Significance of the drop-out time parameter
TRIGger
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...