Instrument Functions
R&S
®
NRP2
99
User Manual 1173.9157.02 ─ 03
Fig. 4-30: List of functions
Impedance measurements "SWR", "RCoeff" and "RLoss"
If you work with the functions for the impedance measurement, i.e. "SWR", "RCoeff" and
"RLoss", note that the incident power must be measured in the primary channel and the
reflected power in the secondary channel.
Continuous Average and Burst Average power measurements
If Continuous Average and Burst Average power measurements have been selected for
the channels that are used as arguments for the measurement function of the "Function
& Unit" dialog box, no additional settings are required to define the measurand. Setting
the unit and defining auxiliaries doesn't affect the measurement itself but tailors the mea-
surement result in such a way that it is convenient for the user.
"Trace & Statistics" mode and "Graph" results window
If the primary channel is operating in the "Trace & Statistics" mode and the results window
is set to "Graph", the envelope power of this channel is automatically shown as a function
of time or in its distribution and distribution density function ("CCDF", "CDF", "PDF"). In
this case, it is not possible to select another measurement function.
Function
The list of functions itemizes all the measurement functions provided by the R&S
NRP2.
Any of the four channels can be assigned primary or secondary channel.
The available units depend on the selected function, see tables
For example, in trace mode, only "Single (A)" and "Ratio (A/B)" are available, and in
statistics mode only one channel is available, which corresponds to "Single (A)".
"Single (A)"
Power in the primary channel
"Ratio (A/B)"
Ratio of the power in the primary channel to the power in the secondary
channel.
"SWR (A,B)"
Standing wave ratio.
"RCoeff (A,B)"
Reflection coefficient.
"RLoss (A,B)"
Return loss.
"Diff (A-B)"
Power difference between the primary and secondary channel.
"Sum (A+B)"
Sum of power in primary and secondary channel.
Configuring measurements
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...