Remote Control - Commands
R&S
®
NRP2
308
User Manual 1173.9157.02 ─ 03
6.13.15 Operation Lower Limit Fail Status Register
The CONDition register contains information as to whether a displayed value is currently
below a configured lower limit and the EVENt register indicates whether a measured
value dropped below a limit value since the last readout of the Operation Lower Limit Fail
Status Register. Details of the behavior are defined by the transition register.
The limit value can be set with the command
CALC:LIM:LOW:DATA
<float_value>
. Command
CALC:LIM:LOW:STAT ON
configures the PTRansition
and NTRansition registers so that the corresponding bit is set in the EVENt register when
the displayed value drops below the limit.
The Operation Lower Limit Fail Status Register can be read by the following commands:
STATus:OPERation:LLIMit[:SUMMary]:CONDition?
STATus:OPERation:LLIMit[:SUMMary][:EVENt]?
Table 6-27: Meaning of bits used in the Operation Lower Limit Fail Status Register
Bit No.
Meaning
0
Not used
1
Window 1 Lower Limit Fail
The measured value returned by the first Calculate block drops below the lower limit value.
2
Window 2 Lower Limit Fail
The measured value returned by the second Calculate block drops below the lower limit value.
3
Window 3 Lower Limit Fail
The measured value returned by the third Calculate block drops below the lower limit value.
4
Window 4 Lower Limit Fail
The measured value returned by the fourth Calculate block drops below the lower limit value.
5
Window 5 Lower Limit Fail
The measured value returned by the fifth Calculate block drops below the lower limit value.
6
Window 6 Lower Limit Fail
The measured value returned by the sixth Calculate block drops below the lower limit value.
7
Window 7 Lower Limit Fail
The measured value returned by the seventh Calculate block drops below the lower limit
value.
8
Window 8 Lower Limit Fail
The measured value returned by the eighth Calculate block drops below the lower limit value.
9 to 14
Not used
15
Bit 15 will never be used.
6.13.16 Operation Upper Limit Fail Status Register
The CONDition register contains information as to whether a displayed value is currently
above a configured upper limit and the EVENt register indicates whether a limit value was
STATus
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...