Remote Control - Commands
R&S
®
NRP2
209
User Manual 1173.9157.02 ─ 03
Example:
Valid command sequence:
FETCh:TSLot:SUM? 577us,8,0,0
FETCh:TSLot:SUM:RELative? 577us,8,0,0
invalid command sequence:
FETCh:BURSt:SUM? 0,0,0
FETCh:TSLot:SUM? 577us,8,0,0
READ:...?
The
READ:...?
command starts a measurement without further configuration and
returns the measurement result. The parameter list is compared to the current settings
before the measurement is started. If the list does not agree, an SCPI error (-221, Setting
conflict) is returned and the command is aborted. Apart from this check,
READ
corre-
sponds to the program message
INIT:IMM; FETCh[1..8]:...?
.
MEASure:...?
The
MEASure:...?
command combines
CONFigure
and
READ
.
Refer to annex
chapter C, "Programming Measurements"
examples to the high level measurement commands.
6.4.2 Basic Commands FETCh? and READ?
FETCh?
The command
FETCh?
without further keywords and parameter list has a special mean-
ing. While omitting parameters of the parameter list normally means that default values
are used and checked, the current measured value under
FETCh?
is supplied when it is
valid. If a measured value is not yet available, processing is suspended until a valid result
is available.
Usage:
Query only
READ?
The
READ?
command without further keywords and parameter list can be compared with
FETCh?
.
READ?
starts a measurement and returns a measured value without checking
the current settings.
Usage:
Query only
6.4.3 Syntactical Structure of High-Level Measurement Commands
The high-level measurement commands can be generally represented as follows:
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...