Remote Control Fundamentals
R&S
®
NRP2
181
User Manual 1173.9157.02 ─ 03
5.8.3 Data Base and Device Hardware
The term "device hardware" refers to that part of the device which performs the mea-
surement function.
The data base contains all parameters required for setting the device hardware. Setting
commands cause a modification of the data set. Before the data is entered in the data
set, it is verified for compatibility both with the other data and with the device hardware.
If the setting is not possible, an error message will be sent to the status reporting system
and the setting will be ignored. After successful completion of the verification, the setting
will immediately be performed.
If the setting influences the result calculation, the
MEAS
uring bit is set in the status oper-
ation register until new measured values are available. The
MEAS
uring bit is described in
more detail in the Status Reporting System section that follows. This bit may be used for
the synchronization of command processing. Queries cause the data-set management
to send the desired data to the output unit.
5.8.4 Status Reporting System
The status reporting system collects information on the device status and makes it avail-
able to the output unit on request. Structure and function of this system are described in
the following section.
5.8.5 Output Unit
The data output unit collects the information requested by the controller from the data set
management. It processes the information in line with the SCPI rules and makes it avail-
able in the output buffer.
5.8.6 Command Sequence and Command Synchronization
All commands are executed in the order in which they are sent. There is no overlapping
of command processing. (Exception: the commands
INITiate
and
CALibrate:ZERO:AUTO ONCE
). The user therefore can determine the sequence
of execution.
If a long program message is terminated by
*OPC
or
*OPC?
, the end of command pro-
cessing is signalled to the controller, provided the controller has been adequately pro-
grammed (see following
Synchronization with *OPC, *OPC? and *WAI
Device Model and Command Processing
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...