Remote Control - Commands
R&S
®
NRP2
263
User Manual 1173.9157.02 ─ 03
Parameters:
<state>
ON | OFF | ONCE
ON
Activates automatic switchover. The sensor always defines a suit-
able filter length which can also be queried via
SENS<[1]..4>:AVER:COUN?
OFF
Deactivates automatic switchover.
ONCE
The filter length is automatically detected during the next mea-
surement and set to that value. Thereafter automatic filter length
detection is switched off. The setting can then be queried via
SENS<[1]..4>:AVER:COUN?
*RST:
depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:MTIMe
<mtime>
If the R&S
NRP2 has to determine the filter length automatically, the filter length can
become large and thus also the measurement time.
An upper time limit can be set via
SENS<[1]..4>:AVER:COUN:AUTO:MTIME
(maxi-
mum time). It should never be exceeded. Undesired long measurement times can thus
be prevented (in the
SENS:AVER:COUN:AUTO:TYPE NSR
mode) if the automatic filter
length switchover is on.
Parameters:
<mtime>
float_value, measurement time
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:RESolution
<resolution>
Defines the number of significant places for linear units and the number of decimal places
for logarithmic units which should be free of noise in the measurement result. This setting
is also performed by the command
on page 243 that tries to set the sensors involved in the measurement results accordingly.
For further details, see
chapter 4, "Instrument Functions"
The setting is only taken into account, if
SENS:AVER:COUN:AUTO:TYPE
is set to
RES
and
SENS:AVER:COUN:AUTO
is
ON
.
However,
SENS<[1]..4>:AVER:COUN:AUTO:RES
does not affect the
DISPlay
com-
mand. The parameters of the two commands are different but have the same meaning.
SENS<[1]..4>:AVER:COUN:AUTO:RES
1
2
3
4
DISP<[1]..4>:RES
1
0.1
0.01
0.001
SENSe (Sensor Commands)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...