Chapter 2 Assembly & Installation
Revision W
HF Series X-ray Generators - Service Manual
2-56
Quantum Medical Imaging
8. Verify the peak voltage at A10TP15 after PREP key has been pressed for at
least 2 seconds (rotor in boost mode) is
greater than
14 V
PEAK
. If necessary,
adjust A10R24 to achieve the correct voltage.
9. Connect a DMM between A10TB1-1 and A10TB1-3.
10. Press the PREP key and verify the voltages as follows:
• For 110 VAC stator tubes (Toshiba E7242X, E7813X, E7252X, E7239X)
96 - 140 VAC during boost, 54 - 66 VAC during run
• For 220 VAC stator tubes (all others tubes not listed above)
183 - 227 VAC during boost, 54 - 66 VAC during run
11. Install cabinet cover.
12. Proceed with system calibration procedure described in Chapter 3,
CALIBRATION.
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...