Appendix D Calibration (Using Generator Service Tools)
Revision W
HF Series X-ray Generators - Service Manual
D-34
Quantum Medical Imaging, LLC
the ground stud on top of the H.V. Transformer (see Figure D-21). Con-
nect the mAs meter in series between removed wires. (
Important: These
wires must be re-connected when finished calibration.
)
Figure D-21. MAS Meter Connection
5. Take an exposure at the setting indicated on the Small Filament Offset
Calibration Screen (i.e., 70 kVp, 75s mA, 15 msec).
If using an oscillo-
scope, set time to 125 msec
.
MAS METER
GREEN/YELLOW
WIRES (2X)
MAS METER
LEADS
HIGH VOLTAGE
TANK A17
(AY20-031 or
AY20-032)
GENERATOR CABINET
(PCB DOOR OPEN)
MAS METER CONNECTION
MAS METER HOOK-UP WIRES
(NORMAL POSITION)
GROUND
STUD
CONNECTOR
A17P1
GREEN/YELLOW
WIRES (2X)
H.V. TANK
(TOP VIEW)
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...