![Quantum ODYSSEY HF Series Скачать руководство пользователя страница 395](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349395.webp)
Appendix B Calibration (DiRex System)
Revision W
HF Series X-ray Generators - Service Manual
B-32
Quantum Medical Imaging, LLC
Filament Offset Calibration
This procedure establishes the filament current offset and gain corrections.
Filament current measurement accuracy is verified using a pre-calibrated
Dynalyzer (or equivalent), non-invasive kV meter, or mAs meter as a bench-
mark. Ensure precision of the measurement instrument is sufficient enough
to assure generator accuracy requirements can be met. The Filament Offset
Calibration procedure may be performed using an oscilloscope, however, it is
not the preferred method. Using this method, the objective is to achieve the
"flattest" possible large and small filament waveforms, with no spikes or
peaks as shown in Figure B-20 below.
Figure B-20. Filament Offset Current Measurement (Using an Oscilloscope)
Before
After
Before
After
Small Filament
Large Filament
FLAT
CURVE, NO
PEAKS
FLAT
CURVE, NO
PEAKS
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...