Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
5-7
READY
LED not
illuminated
1. Generator Cabinet ON/OFF Cir-
cuit Breakre is OFF.
2. ON/STDBY Rocker Switch on
OCP (non-Q-Connect Sys-
tems) or Power/Expose Control
Box (Q-Connect Systems) is set
to STDBY.
3. Facility electrical has been
interrupted.
4. Filament automatically shuts off
after 60 minutes of inactivity.
4. The exposure setting(s) exceed
the tube or model limits.
5. Faulty communications
between generator and OCP.
6. Incorrect generator program or
generator is not working.
7. Lost communication between
OCP and generator cabinet.
1.Verify that the ON/OFF Circuit
Breaker on Generator is ON.
2. Set ON/STDBY Rocker Switch on
OCP (non-Q-Connect Systems) or
Power/Expose Control Box (Q-
Connect Systems) to ON.
3. Check, and if possible, reset Facil-
ity electrical service.
4. Press PREP key. An error mes-
sage displays on the operator con-
trol panel. Refer to appropriate
error message listed in this table
for error details. Press any key to
clear error. System should now be
ready for exposure.
5. Non-Q-Connect Systems only:
Check serial cable connection
between OCP and generator cabi-
net and jumper settings on OCP
Control Board (AY40-004S1).
400 Volt
Anode
See "Line/400V Anode" message
See "Line/400V Anode" message
400 Volt
Cathode
See "Line/400V Cathode" mes-
sage
See "Line/400V Cathode" message
AC Volt
Diff
Output by Logic board, indicating
a hardware failure when a signifi-
cant difference between Anode
and Cathode voltage is detected.
Note: This error applicable on Q-
Connect Systems only.
Proceed to "KVP Error" and trouble-
shoot accordingly.
ADC
Negative
15
See "Minus 15V" message
See "Minus 15V" message
ADC
Negative
Reference
See "Vref-" message
See "Vref-" message
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...