Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-19
CAUTION! When routing the High Voltage Cables
into the generator cabinet through the access
openings, the anode and cathode cable must each
be routed separate from all other system cables,
except the Tube Rotor Cable, as shown in Figure
2-13. Doing so will reduce the possibility of elec-
tromagnetic noise coupling and subsequent deg-
radation of system performance.
Figure 2-13. Routing High Voltage Cables into the Generator Cabinet
ROTOR
CABLE
ALL OTHER
CABLES
GENERATOR CABINET (VIEWED FROM REAR)
HIGH
VOLTAGE
CABLES
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...