Chapter 3 Calibration
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
3-19
1. Select Tomo from the Service Configuration Menu (see Figure 3-4). The
system displays the Tomo menu (see Figure 3-13).
Figure 3-13. Setting Tomography System Interface Configuration
2. Press the key next to the Tomo field until the desired tomography sys-
tem interface setting (Siemens, Pausch, or OFF (none)) is displayed. If
no tomography option is interfaced, set to OFF. Otherwise, set to either
Siemens or Pausch as required.
3. Press the key next to Tomo Times field. Depending on whether "Sie-
mens" or "Pausch" tomo was selected, the appropriate Tomo Times
screen (see Figure 3-14 or 3-15) is displayed:
Figure 3-14. Tomo Times Screen - Siemens Tomo Selected
PRESS THIS
KEY TO SET
TOMO TIME
SELECTIONS
PRESS THIS
KEY TO SET
TOMO MODE
ON SIEMENS
TOMO SYSTEMS,
DO NOT
CHANGE ANY
TOMO TIME
SETTINGS
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...