Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
5-79
a. Install three (3) OCP Interface Board A1A1 mounting stand-
offs on Logic Board A1 in same locations from which they
were removed.
b. Place the OCP Interface Board A1A1 on mounting standoffs
and secure using one (1) screw and two (2) hex nuts.
c. Connect ethernet cables to the OCP Interface Board A1A1
connectors J1 and J5.
3. Connect all cables to Logic Board A1. Remove tags.
Note: When installing IC's, ensure that pin one on the
chip is aligned correctly with pin 1 on the IC socket
mounted on the board. Do not use software label to
orient IC!
4. Insert Logic Master IC U9 (marked "QLBM") into U9 socket on
Logic Board A1.
5. Insert the Logic Slave IC U14 (marked "QLBS") into U14 socket
on Logic Board A1.
6. On Logic Boards revision G and higher, verify jumpers JP1, JP6,
and JP7 are configured as follows:
a. Jumper JP1 - must always be installed
b. Jumper JP6 - installed only if system is equipped with high-
speed starter (QG-HSS) option. Do not install on standard,
single-speed rotor systems (i.e., systems equipped with Rotor
Drive Board A10 AY40-013T).
c. Jumper JP7 - install jumper on pins 2-3
7. Set the ON/OFF circuit breaker CB1 on the generator cabinet to
ON position.
8. Set the power on/standby switch on the Operator Control Panel
(OCP) to the on ( ) position.
9. Verify generator passes startup self-tests.
10. Access Service Mode, Service Configuration and Settings screens,
and verify generator configuration settings are correct.
11. Access Service Mode, AD Calibration screen. Verify all voltages
are within tolerance. If not, calibrate in accordance with "A/D
Calibration Procedure" in Chapter 3 (ODYSSEY) or Appendix B
(DiRex Integration generators), or Appendix D (all other inte-
grated generators).
12. Spot check kVp and mA calibration and, if needed, re-calibrate
system.
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...