![Quantum ODYSSEY HF Series Скачать руководство пользователя страница 195](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349195.webp)
Chapter 5 Service Instructions
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
5-11
Anode Volt
(Cont’d)
6. Verify AC line voltage meets speci-
fication during an exposure: Check
capacitor voltage (C9-C12 and
C13-C16 on single-phase units,
C1-C4 on SE units, or C9-C12 and
C13-C16 on three-phase units) is
b350 and +440 VDC.
The power must be sufficient to
support the exposure (i.e., check
for weak AC Line).
7. Verify that all grounds are good
(JP2-JP9 installed or shorted on
Logic Board A1 (AY40-006S)).
APR Tree
Load Error
{text
string}
An error occurred while loading
the APR tree. Note: {text string}
will contain additional details.
Call Technical Service/Support.
Backup
Backup time was reached based
on current technique factors.
1. Reduce technique factor(s). Re-
take exposure.
2. Verify ion chamber cable connec-
tions are correct and secure.
3. On AEC Control Board A11
(AY40-031S or AY40-027S), check
+15 VDC and -15 VDC outputs are
not loaded down. If so, remove ion
chamber cable connection and re-
check voltages. If +15 VDC and /or
-15 VDC outputs are now good, ion
chamber is at fault. If still loaded
down, replace AEC Control Board.
4. Verify the Backup Time/mAs set-
tings in APR view (if caused during
APR mode) and/or Backup Time/
mAs in Service Settings configura-
tion screen are set correctly (i.e.,
not too low).
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...