Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-64
Quantum Medical Imaging, LLC
DiRex Integrated System - Statistics Screen
The Statistics Screen is accessed as follows:
1. Set the ON/OFF circuit breaker CB1 on the generator cabinet to
ON position.
2. Set the power on/standby switch on the Q-Connect Control Panel
(OCP) to the standby ( ) position, then to the on ( ) position.
The green power LED lamp should illuminate.
3. From the Main OCP Overlay Menu, double click the generator sta-
tus icon. The "Enter Password" screen appears.
4. In the Enter Password screen, enter the default Service Techni-
cian password (58623) using the keypad, then press Enter.
5. Upon validation of the password, the Service Options Menu dis-
plays. Press the Generator Service button.
Figure 5-5i. Service Options Menu
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...