![Quantum ODYSSEY HF Series Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349121.webp)
Chapter 3 Calibration
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging, LLC
3-29
SYSTEM CALIBRATION
CAUTION! Before calibrating unit, perform warm-up
procedures recommended by tube manufacturer or
the warm-up procedure described in Chapter 3 of the
Operator’s Manual.
Important! On generators that will be interfaced to a
tomography system; Set Tomo Mode to "Off" during
prior to performing generator calibration, then set
Tomo Mode back to "On" following calibration.
To access the system calibration functions, press the key next to the Calibration
field in the Service Menu (see Figure 3-3). The Service Calibration menu displays,
shown below:
Figure 3-24. Service Calibration Menu
The Service Calibration menu provides access to the system's various calibration
functions. The generator was pre-calibrated at the factory, however, the system
will need to be re-calibrated to assure proper operation, verify compliance and set
AEC optical density. These procedures should be performed in the order pre-
sented in this chapter. The following is a brief description of each of the Service
Calibration Menu selections:
•
Factory Reset: Accesses the Factory Reset Menu, which provides a list
of resettable system calibration and configuration parameters including:
» Apr (user-customized APR procedures)
» Tube curves
» AEC (AEC calibration values)
» Sentence (Modified Sentences)
» History (Error Log)
» Objects (all other of the program variables and subroutines
excluding the above parameters; includes kV, mA, and filament
calibration values)
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...