![Quantum ODYSSEY HF Series Скачать руководство пользователя страница 230](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349230.webp)
Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-46
Quantum Medical Imaging, LLC
Wall
Bucky
The wall bucky has been selected
but did not return the READY sig-
nal to start exposure. (This may
take up to five seconds after expo-
sure is requested.)
1. Verify the following did not occur:
The "WALL" receptor key on the
OCP was selected on a system not
equipped with a wall receptor.
2. Check the bucky cable connec-
tions at both ends (i.e., Interface
Board A9 (AY40-023T or AY40-
034T) terminal block TB4 and the
table bucky). Refer to bucky inter-
connection diagrams contained in
Chapter 6, DIAGRAMS for proper
wire connections.
3. On Interface Board A9 (AY40-023T
or AY40-034T), verify 120 VAC
bucky power fuse F12 (2A, 250V
slow-blow) is not open.
4. On Interface Board A9, check if
switch SW1 is in correct position
("M" position if wall stand is
equipped with Midwest bucky, non-
"M" position for all other type buck-
ies).
5.Verify fuse in bucky is not open.
Recycle bucky power by turning
generator off, then on. If error still
occurs, replace bucky control
board.
6. On Logic Board A1 (AY40-006S),
check for an active low signal at
J4-5 during expose. If not present,
replace Logic Board (including
firmware). If present, check cable
connection between Interface
Board A9J6 and Logic Board
A1J4. If connection is good,
replace Interface Board.
Warning -
Using
default
APR tree
Occurs in conjunction with "APR
Tree Load Error".
Because of a load error the default
APR tree will be used. This WILL
NOT overwrite the current APR tree if
it exists. You will get this message
until you fix the tree. Load APR editor
and fix the tree.
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...