Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-13
For non-SE generators, connect AC input cable as follows:
1. Mount cable clamp (included with system) to the hole on the lower right
or left side of generator cabinet chassis (see Figure 2-7).
Figure 2-7. Line Input Cable Clamp Installation
2. Route the AC input cable through cable clamp to Terminal Block TB1 and
tighten cable clamp.
3. On single-phase systems:
a. Connect ground wire to ground lug mounted on rear wall of genera-
tor chassis, to the left of terminal block TB1 (see Figure 2-8).
Figure 2-8. Single-Phase Systems - Ground Lug and Terminal
Block TB1 Locations
GENERATOR
CHASSIS
LINE INPUT
CABLE CLAMP
(TYPICAL)
TERMINAL
BLOCK TB1
GROUND
LUG
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...