Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-27
Tube Rotor Cable (Systems with AID External High Speed Starter
QG-HSS Option Only)
Refer to Figure 2-18, the AID High Speed Starter installation manual and/or
the appropriate interconnection diagram in Chapter 6, DIAGRAMS for high
speed starter connections.
Note: This procedure applies only to sys-
tems equipped with the externally-connected AID High Speed
Starter option.
1. Connect one end of X-ray Tube Rotor Cable AY51-032T to the appropriate
connector/terminals on AID High Speed Starter LC102 terminal block TS2
(see Figure 2-18).
2. Connect the opposite end to the X-ray tube’s terminal strip.
3. Route the X-ray Tube Rotor Cable AY51-033 (Toshiba tubes) through the
middle cable access opening located in the top rear wall of the generator
chassis.
4. Connect one end of the High Speed Starter Cable AY51-033 to the appro-
priate connector/terminals on AID High Speed Starter terminal strips TB1
and TB2 and connect to Cable Assembly AY51-032T (temperature sense
connector).
5. Connect the opposite end of the High Speed Starter Cable AY51-033 to
connector J8 on Logic Board A1, AY40-006S.
6. Connect 220 VAC power input to High Speed Starter.
7. The AID High Speed Starter was factory pre-configured for the x-ray tube
supplied with the generator. Verify boost and brake settings are correct
according tube manufacturer’s data sheets (refer to Table 2-3). Verify the
high speed starter is configured with the correct capacitors for the tube
used with the system: Toshiba E7252X and E7813X should be configured
for 45 uF. All other tube types require 30 uF.
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...