Appendix B Calibration (DiRex System)
Revision W
HF Series X-ray Generators - Service Manual
B-28
Quantum Medical Imaging, LLC
1. In the Calibration Menu, press AD button (see Figure B-16). The AD Cal-
ibration Menu appears:
Figure B-17. AD Calibration Menu
2. Refer to Table B-3 and connect a voltmeter to the test point correspond-
ing with the voltage level selected for calibration.
3. Select a voltage level to calibrate by pressing its corresponding Calibrate
button (choices are Cathode (400V), Anode (400V), +15 VDC, -15 VDC,
Reg24 VDC, Unreg24 VDC, and +5 VDC). For this
example, "Cathode" is selected.
PRESS TO SCROLL UP/
DOWN LIST FOR
ADDITIONAL SETTINGS
VOLTAGE
SELECTIONS
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...