![Quantum ODYSSEY HF Series Скачать руководство пользователя страница 190](http://html1.mh-extra.com/html/quantum/odyssey-hf-series/odyssey-hf-series_service-manual_3538349190.webp)
Chapter 5 Service Instructions
Revision W
HF Series X-ray Generators - Service Manual
5-6
Quantum Medical Imaging, LLC
code displayed on the Operator Control Panel:
Table 5-1. Error Messages and Possible Cause(s)
Error
Message
Possible Cause(s)
Remedial Action
READY
LED blink-
ing
Note: A brief period of blinking is
normal when the unit is first turned
on or following an exposure.
1. 400 V Not ready for exposure.
2. On stored energy (SE) units,
batteries not ready for exposure
(i.e., in charging mode).
3. Heat units are exceeded.
1. If blinking is continuous, verify the
line voltage is high enough. (Check
voltage across capacitor bank (C9-
C12 and C13-C16 on single-phase
units, C1-C4 on SE units, or C9-
C12 and C13-C16 on three-phase
units) to see if it is out of range
(i.e., 380-420 VDC).)
2. Check AC input transformer tap
adjustment for correct line voltage
tapping.
3. Check for 24 VDC at Power Sup-
ply Board A6 (AY40-005T) +24
VDC test point. If +24 VDC not
present, turn off generator and
remove plug from Power Supply
Board A6 (AY40-005T) connector
J2 and re-check +24 VDC test
point. If +24 VDC is present,
something is loading down voltage.
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...